EIGER R for Laboratory

Specification
EIGER R 1M
EIGER R 4M
Number of detector modules
1 x 2
2 x 4
Sensitive area (width x height) [mm2]
77.2 x 79.9
155.2 x 162.5
Pixel size [μm2]
75 x 75
75 x 75
Total number of pixels
1030 x 1065 = 1,096,950
2070 x 2167 = 4,485,690
Gap width, vertical [pixel]
-/37
10/37
Inactive area [%]
3.5
5.6
Defective pixels [%]
< 0.03
< 0.03
Maximum frame rate [Hz]
10
5
Readout time
continuous readout, 3 μs dead time, duty cycle >99.6%
continuous readout, 3 μs dead time, duty cycle >99.6%
Point-spread function
1 pixel
1 pixel
Sensor thickness [μm]
450
450
Threshold energy [keV]
2.7 - 18
2.7 - 18
Maximum count rate [phts/s/mm2]
5 · 108
5 · 108
Image bit depth [bit]
32
32
Dimensions (WHD) [mm3]
114 x 133 x 240
235 x 235 x 372
Weight [kg]
3.9
15
Power consumption [W]
75
300