PILATUS3 R CdTe 100K-M

The PILATUS3 R CdTe 100K-M brings the advantages of HPC technology and direct detection in CdTe to high-energy applications in the laboratory. The system with an active area of
83.8 x 33.5 mm2 outclasses many other detectors for laboratory XRD with hard radiation. Capture more of your precious photons from Mo, Ag, or In sources and detect them with greater than 90% quantum efficiency. When speed and data-collection efficiency matters for you in PDF, in-situ, or in-operando measurements, the PILATUS3 R CdTe 100K-M is the best choice for you.

In addition, the 100K-M offers all the features of the PILATUS3 X CdTe hybrid photon counting detectors, such as noise-free detection for highest dynamic range and optimal point spread function.

Furthermore, you can exploit the optional vacuum compatibility (down to 10-6 mbar) of the detector for complete elimination of air scatter and absorption. Electrical and cooling feedthrough sets are available as accessories and make integration of the 100K-M into your vacuum chamber easy.

Specifications PILATUS3 R CdTe 100K-M

Number of detector modules (W x H) 1 x 1
Sensitive area (width x height) [mm²] 83.8 x 33.5
Pixel size [µm²] 172 x 172
Total number of pixels (horiz. x vert.) 487 x 195 = 94,965
Gap width vertical / horizontal [pixels] 1 pixel wide
Gap between modules (horiz./vert.), *plus 1 pixels horizontal gap on each module [pixel] -* / -
Inactive area [%] 0.2
Defective pixels [%] < 0.1
Maximum frame rate [Hz] 20
Readout time [ms] 7
Point-spread function 1 pixel (FWHM)
Threshold energy [keV] 8 - 40
Maximum count rate [phts/s/pixel] 1 x 10⁷
Counter depth [bit] 20 bits (1,048,576 counts)
Power consumption [W] 30
Dimensions (W x H x D) [mm³] 114 x 69 x 118 (Detector head)
Weight [kg] 0.9 (Detector head)
Cooling Water-cooled (Detector head)
Electronics cooling Air-cooled

 

All specifications are subject to change without notice.